Tui tlak Angle Meter Calibration chungchang sawifiahna
Aug 19, 2025
Water Drop Angle Meter calibration specifications hi instrument tehna dik leh inmilna atana thil pawimawh tak a ni. Calibration process-ah hian a tlangpuiin a hnuaia step te hi a tel thin: 1. Standard Sample buatsaih dan: Calibration hi standard sample contact angle hriat lar tak hmanga tih a ni (polytetrafluoroethylene plate contact angle 110℃vel nei ang chi). 2. Instrument Leveling: Test platform chu level a nih ngei ngei tur a ni a, spirit level hmangin level chu Less than emaw a tlukpui emaw a dik taka siamrem tur a ni 0.1℃. 3. Software Calibration: Software-a hman tur built-in calibration module to adjust the image acquisition system’s focus and light source intensity to ensure a clear water droplet outline. 4. Repeatability Testing: Standard sample khatah tehna tam tak (5 aia tam emaw a tlukpui emaw a rawt) ti la, standard deviation 1℃aia tlem emaw, a tlukpui emaw chhut rawh . 5. Environmental Control: Calibration hi tih tur a ni temperature (23±2℃) leh humidity awm reng a ni (50±5%RH) awmna hmun a ni. Thla thum danah emaw, key components thlak hnuah emaw calibration neih a tha.
